Product Operation Scope
The fingertip Oximeter can be used to measure human Haemoglobin Saturation and heart rate through finger. The product applies to using in family, hospital (including clinical use in internist/surgery, Anaesthesia, paediatrics, intensive care and etc.), Oxygen Club, social medical organizations, physical care in sports (It can be used before or after sports. Operation in sport procedure is not recommended). It is also applicable to enthusiasts on mountaineering, patients (convalescents at home or those need first aid treatment), elders over 60, those work more than 12 hours, sporters and those work on the hermetic circumstance, etc. The product is not suitable to monitor patient continuously.
Detailed descriptions of product functions:
1. Display Type: LCD display or OLED display
2. SpO2
Measurement range70%99%
Accuracy±2% on the stage of 70%99%,Unspecified(≤70%) for SPO2
Resolution±1%
Low Perfusion≤0.4%
3. PR
Measurement range30BPM240BPM
Accuracy±1BPM±1%the larger one
4. Power two AAA 1.5V alkaline batteries
5. Power consumptionbelow 30mA
6. Automatic power-offthe product shuts off by itself when no finger is in the product in 8 seconds
7. Dimension
8. Operation Environment
Operation Temperature540
Storage Temperature-1040
Ambient Humidity15%80% on operation
10%80% in storage
Air Pressure: 70kPa106kPa
9. Declaration: EMC of this product comply with IEC60601-1-2 standard.
10. Measurement Performance in Low Perfusion Condition: required test equipment (BIO-TEK INDEX Pulse Oximeter tester) could measure the available pulse wave with the amplitude of 6% of the simulation pulse wave amplitude.
11. Interference Resistance Capacity against Ambient Light: Device works normally when mixed noise produced by BIO-TEK INDEX Pulse Oximeter tester